Equipment > Audiometers > Madsen Xeta

Madsen Xeta

Madsen Xeta

Intuitive and well thought out, the MADSEN™ Xeta™ audiometer from Otometrics is perfect for those who want fast, efficient testing. Air, bone, and masking, combined with valuable features such as an intuitive layout, flexible storage, and a built-in PC connection, ensure efficient audiometry - in the clinic or in the field

Features: 
Key features
Air, bone and masking
Pure, warble and pulsed tones
Auto threshold and auto screening
Special tests (optional): ABLB, SISI, Stenger
Built-in Talk-over communication
Patient data storage (50 planned patients and 75 full audiograms)
User configurable setups
Connects with OTOsuite™
Audiometry software (NOAH compatible)
Masking Assistant™
Technical specifications
Pure tone frequencies
AC: 125-8000 Hz
BC: 250-8000 Hz
Masking: Narrow Band Noise
Stimulus levels
AC: -10 to 120 dB HL at mid-frequencies, 5 dB steps
BC: -10 to 70 dB HL at mid-frequencies, 5 dB steps
Automatic threshold test
Test adheres to the modified (ascending) Hughson & Westlake method
Normal and Quick mode
Automatic screening test
Optional threshold search for failed frequencies
PC interface
RS232 (USB converter available)
Dimensions and Weight
Approx. 355x415x130 mm, 14x16.5x5.1 inches Approx. 4 kg, 8,8 lb.
Power supply
Internal, 100-240 VAC, 50/60 Hz
Standards
Audiometer: EN 60645-1 and ANSI S3.6 Patient safety: Complies with EN 60601-1, Class 1,
Type B: U2601-1; CAN/CSA-C22.2 NO 601.1-90
EMC: EN 60601-1-2
Accessories
Standard
Choice of: TDH39, E-A-RTONE 3A insert phone, ME70 or HOLMCO
Patient responder switch
Optional
Bone conductor
E-A-RTONE 3A Mono insert masking phone
OTOsuite™
Audiometry software
Carrying case
Special test